DMC-3 (Ex-Zone 2)
Robust piezo-resistive Δp-SENSOR-TRANSMITTER industrial-grade with good long-term stability and a high overpressure safety for use in Ex-Zone 2. 3 linear Outputs: 0-10 V, 0-20 mA and 4-20 mA are part of the standard version. The TRANSMITTERs are available without or with 3-digit digital display.
Examples: air conditioning and ventilation systems, medical applications, clean room technology, filter control, level control, flow control, industry, etc.
For air conditioning or ventilation applications with the almost always present turbulences, the integrated adjustable electronic Δp-damping of the outputs and the display represents a further advantage.
Features
Differential pressure, relative pressure (positive or negative pressure)
Measuring ranges: 0 - 0.5 mbar to 0 - 10 bar and ±0.5 mbar to ±1000 mbar available
available with and without digital display
Pressure sensor highly overloadable
Pressure sensor temperature compensated
Integrated filters in the pressure connections
Adjustable Δp-damping as standard: 1s, 10s, 20s, 30s
Extreme ambient temperature range -20 °C to +40 °C (4°F to +108°F) according to ATEX
3 analogue outputs as standard: 0-10 V, 0-20 mA and 4-20 mA, linear, short-circuit-proof
Available for supply voltage 24 VDC, 12/24 V, 50-60 Hz and 115-120/230-240 V, 50-60 Hz
Housing protection class IP 65, aluminium
Marking: Ex II 3G Ex nR IIC T6 Gc X or Ex II 3G Ex nA IIC T6 Gc, depending on version
Documentation